Drawer module 19" / 4 HE
The device is equipped with an electronic
high voltage source.
For the adaptation of test equipment plugs for
the following units are implemented on the rear:
- PE test probe
- HV test probe (1)
- 7pole system plug
System-Interfaces:
Ethernet (alternative RS232C or USB)
Digital-Interface:
Inputs:
- Start, Stop Remote
Outputs:
- OK, I>Imax, System failure, HV On,
Test voltage:
0,1..2,5/0,2..5 kV AC
Voltage programmable
Floating output
Tripping current:
0 .. 10/100mA AC
Measuring accuracy:
Voltage measurement:
1% of rdg. ± 3 digit
Resolution voltage measurement:
10 V
current measurement (apparent /active c.):
100 mA - range:
1 % of rdg. ± 3 digit
10 mA - range:
1 % of rdg. ± 3 digit
Resolution current measurement:
100 mA - range:
100 uA
10 mA - range:
10 uA
Programmable test time:
0,5 .. 999,9 s
PE conductor measurement
Open-circuit voltage:
6 V / 12 V
Test current:
10 ... 32 A
Resistance measuring range:
0 .. 1,2 Ohm
Resolution resistance measurement:
1 mOhm
Voltage drop measuring range:
0 .. 12 V
Resolution voltage measurement:
10 mV
Measurement range current measurement:
0..50 A
Resolution current measurement:
100 mA
Four-wire method
Accuracy:
1,5% of rdg. ± 3 digit
Programmable test time:
0,5 .. 999,9 s
Insulation resistance measurement
Test voltage:
50 .. 1200 VDC
Measurement ranges:
0,1 .. 10/100/1000 MOhm
Resolution resistance measurement:
10,0 MOhm range:
100 kOhm
100 MOhm - range:
1 MOhm
1,00 GOhm range:
10 MOhm
Accuracy:
1% of rdg. ± 3 digit
Programmable test time:
0,5 .. 999,9 s
Programmable down time:
0,5 .. 99,9 s
Display / Operating unit:
4,3" LC touch display; single colour
Interface:
Ethernet (Alternative RS232C or USB)
Test run :
fully automatic / manually
AC power supply:
230V ± 10% / 49 .. 61Hz
Incl. ELABO-Calibration-Certificate.
Weight: appr. 24 kg
W = 483 mm; H = 177 mm; D = 360 mm