Test voltage:
50 ... 1200V DC
Measuring range:
100k - 10GOhm
Resistance measuring range:
10kOhm ... 5GOhm
The test current is limitted to
1mA. Therefore further safety precautions
for the operator are not necessary.
The control and parameterization is done via the
built-in Ethernet interface.
Optional are the connections for the new
ELABO remote control (15 - pin D-SUB,
female) and the digital interface (25-
pole D-SUB, female).
The interfaces and the power connection are on
the lower side accessible.
The mains connection is as GST 18 built-in
connector executed.
Inclusive ELABO calibration certificate
AC power supply:
100V - 240V, 50Hz / 60Hz +24 Vdc external (DIO, RC)
Test voltage:
50V ... 1200V, variable adjustable in 1V increments
Output power:
max. 1W
Measuring range:
100kOhm ... 10GOhm
Monitoring:
R_min and R_max (contact monitoring) U_set (test voltage)
Adjustable test time:
0.5 s ... 1d adjustable in 0.1s increments
Automatic range selection or fixed
Measuring range after worst-case analysis
Display of measured values, deviations,
Accuracy:
100,0 kOhm … 999,9 kOhm: 100 Ohm
1,000 MOhm … 9,999 MOhm: 1 kOhm
10,00 MOhm … 99,99 MOhm: 10 kOhm
100,0 MOhm … 999,9 MOhm: 100 kOhm
1,000 GOhm … 9,999 MOhm: 1 MOhm
ab 10,00 GOhm: 10 MOhm
Voltage:
100,0 kOhm … 999,9 kOhm: ±0,5% v. MW, ±2 Digit
100,0 MOhm … 999,9 MOhm: ±1,0% v. MW, ±2 Digit
1,000 GOhm … 4,999 GOhm: ±2,0% v. MW, ±2 Digit
5,000 GOhm … 9,999 GOhm: ±3,0% v. MW, ±2 Digit
>10,00 GOhm: ±5,0% v. MW, ±2 Digit
Display resolution:
0,05kV ... 1,2kV: 0,01kV
Measuring accuracy:
up to 1,2kV: ± 0.5%, ± 1Digit
Including ELABO extension module DIO
4 digital inputs and 6 digital outputs
Additional 4 digital inputs:
- Start
- Stop
- Remote
- Reserve
6 digital outputs:
- Faulty test object
- Test running
- Test OK
- Pulse (250ms) to open a pneum. test cover
- System error
- Control ext. discharge relay
Dimensions:
Width = 61mm (12TE)
Depth = 235 mm
Height = 220 mm
Technical modifications and errors excepted.
Optionally available with calibration.