Installation in switch cabinets
Internal and external supply of the DUT possible
Testing in accordance with standards in conjunction with elution software + standards licenses
Max. 2 leakage current basic measuring circuits can be integrated (options)
(MD according to EN60990 B3/B4/B5, MD according to EN60601 MED B12, MD according to UL/CSA)
Control and parameterization via Ethernet interface and protocol
Testing in various operating modes and fault scenarios via basic test steps, predefined in standards files
(normal operation without fault, normal operation with N fault, normal operation with PE fault, reverse polarity operation without fault, reverse polarity operation
with N-fault, reverse polarity operation with PE-fault)
Variable test sequence, execution of the test (basic test steps) can be individually selected and deselected.
Adjustable test time and warm-up phases before performing the test
Front panel connections (terminal blocks):
External power supply (DUT)
Test object connection, 1-phase, vi1, vi2
Additional connections for tests according to
EN60601-1, MED
Connections on the underside:
Mains connection (GST 18 chassis connector)
External power supply 24VDC, max. 1A
(fused) for digital I/O, RC, warning lamp
(2-pin WAGO plug)
Ethernet interface (RJ45 socket)
Connection for warning lamp set 24VDC (M12 socket, 5-pin)
RC remote control connection with display (15-pin D-SUB, female)
Digital I/O connection (25-pin D-SUB, female)
Connections for 1-phase supply module
for regulating the operating / test voltage
(plug connector (3-pin) with interlock for power supply and
power supply and M12 socket (8-pin) for regulating the
control of the electromechanical transformer)
T E C H N I C A L E D A T E S :
Supply voltages (test module):
Mains connection, 230V AC, 1-phase, max. 2A
(fused), 50/60Hz, max.
Mains voltage fluctuation ±10 %
External power supply 24 V DC, max. 1A
(fused), (DIO, RC, warning lamp, SK)
Supply voltage / test voltage (DUT):
Internal supply via separate supply module
supply module with earth potential reference,
U_net ≤ 280V AC, 1-phase, max. 15A (rev.
fused), 50/60Hz mains-dependent
External power supply, 50V ≤ 280V AC, 1-
phase, max. 15A (rev. fused), 50 ≤ 400Hz
Measurements during operation / testing of the device under test:
Operating / test voltage: 10V ... 285V AC, 50Hz ≤ 400Hz
Operating current: 0.1A ... 15A AC, 1-phase
Measurements leakage current: 5 µA... 10 mA DC / AC, 50Hz ≤ 1MHz
3 leakage current measuring ranges: 0.1mA / 1mA / 10mA
Measurement types: I_trms, I_ac, I_dc, I_peak,
I_mad (depending on the test standard used)
Fixed measuring range according to I_lc_max or
automatic measuring range adjustment
Monitoring:
Operating voltage: U_op_min ≤ U_op ≤ (U_op+ 10%)
Test voltage: U_test_min ≤ U_test ≤ (U_test + 10%)
Operating current: I_op_min ≤ I_op ≤ I_op_max
Leakage current: I_lc_min ≤ I_lc ≤ I_lc_max (max. 2 leakage currents for tests according to standard EN60601-1, MED)
Adjustable test time and warm-up phases:
1s ? 1d adjustable in 0.1s steps
Setting accuracy, deviations:
Operating / test voltage: U_netz ≤280.0V
Measuring accuracy, deviations, display of
measured values (resolution): (website)
Including ELABO expansion module Digital I/O:
4 digital inputs: START, STOP, REMOTE, reserve
6 digital outputs: test item iO, test item
niO, system error, test in progress, pulse (250ms)
for pneumatic test cover, ext. discharge via additional relay
Mounting: 6 holes for screwing to a mounting plate
Dimensions: 426mm (84TE)x 220mm x 235mm (W/H/D)
Weight: approx. 8.5kg
Subject to technical changes.